Item
|
value
|
|
Warranty
|
1 Year
|
|
Place of Origin
|
Japan
|
|
Brand Name
|
YOKOGAWA
|
|
Model Number
|
GC8000
|
|
Product name
|
Process Gas Chromatograph
|
|
Measurable objects
|
Gases or volatile liquids (400°C or lower boiling point)
|
|
Analysis method
|
Gas chromatography
|
|
Measurable range
|
TCD: 1 ppm to 100%
MTCD: 1 ppm to 100% FID: 1 ppm to 100% FID with methanizer: 1 ppm to 0.1% FPD: 1 ppm to 0.1% |
|
Number of components to be measured
|
Maximum of 999 (total number of components in all streams including calibration standard sample stream)
|
|
Analysis period
|
Maximum of 21600.0 seconds (six hours)
|
|
Number of streams to be measured
|
Maximum of 31 (including calibration standard sample stream)
|
|
Material of sample-contact parts
|
RV: 316SS, Hastelloy-C, Rulon, PTFE (Teflon, Bearee)
LSV: 316SS, Hastelloy-C, Rulon, PTFE (Teflon, Bearee), Glass, Fluororubber (Viton), perfloroelastomer (Kalrez) |
|
Repeatability
|
Depends on analysis conditions. Gas sample: ±1% of full scale for measuring ranges (2σ) Liquid sample: ±2% of full scale for measuring ranges (2σ) |
|
Type of protection
|
Pressurized enclosure, flame proof enclosure, increased safety, intrinsic safety and type of protection
Certification standard: TIIS, FM, CSA, ATEX (DEKRA), IECEx (DEKRA), NEPSI TIIS: Ex pd IIB+H2 T1 to T4 FM/CSA: Type X Pressurization and Explosionproof for Class I, Division 1, Groups B, C and D. T1 to T4 Type X and Y Pressurization for Class I, Division 1, Groups B, C and D. T1 to T4 ATEX: II2G Ex db pxb IIB+H2 T1, T2, T3 or T4 Gb II2G Ex db ec ic nC pxb pyb IIB+H2 T3 or T4 Gb IECEx: Ex db pxb IIB+H2 T1, T2, T3 or T4 Gb Ex db ec ic nC pxb pyb IIB+H2 T3 or T4 Gb NEPSI: Ex db pxb IIB+H2 T1, T2, T3 or T4 Gb Ex db ec ic nC pxb pyb IIB+H2 T3 or T4 Gb |
|
|
||
|
Protection degree of enclosure
|
Type 3R (ANSI/UL 50E, CSA C22.2 No. 94.2) 、Equivalent to IP54 (IEC 60529) (dust and water resistant structure)
|
|
|
||
|
Ambient condition during operation
|
-10 to 50°C, 95 % RH or less (no condensation)
Depends on analysis conditions. |
|
|
||
|
Weight
|
|
Wall-mounting version
|
Self-standing version
|
||
|
|
Type 1
|
Approx. 100 kg
|
Approx. 140 kg
|
||
|
|
Type 2
|
Approx. 155 kg
|
Approx. 190 kg
|
||
|
|
Type 3
|
Approx. 200 kg
|
Approx. 220 kg
|
||
|
|
Type 4
|
Approx. 140 kg
|
Approx. 170 kg
|
||
|
|
Type 5
|
Approx. 175 kg
|
Approx. 195 kg
|
||
|
|
Type 6Approx.
|
85 kg
|
-
|
||
|
Volume
|
Large isothermal oven: Approx 45 L, Isothermal oven: Approx 31 L
|
|
Oven temperature range
|
Type1 to 5 : 55 to 225°C Type 6 : 55 to 135°C (Temperature can be set in one-degree step.) |
|
Temperature stability
|
±0.03°C* *Reference value
|
|
Temperature control
|
PID |
|
Analog Input/Output
|
Maximum of 16/Maximum of 32
|
|
Contact Input/Output
|
Maximum of 32/Maximum of 20
|
|
PC communication
|
Ethernet communication
|
|
DCS communication
|
RS422 Protocol: MODBUS, Y-Protocol, BTU for Japan Ethernet communication Protocol: Modbus/TCP |
|
Power supply
|
Type 1, 2, 3, 5, 6:
100/110/115/120/200/220/230/240 V AC, ±10%, 50/60 Hz ±5% Type 4: 200/220/230/240 V AC, ±10%, 50/60 Hz ±5% Note: Depends on analysis conditions. Protection devices, such as a breaker, are required to avoid overcurrent. |
|
Maximum rated power
|
Type 1: 0.8 to 1.6 kVA, Type 2: 1.4 to 2.9 kVA, Type 3: 2.0 to 4.3 kVA, Type 4: 1.7 to 3.7 kVA Type 5: 1.9 to 3.0 kVA, Type 6: 0.8 to 1.2 kVA |
|
Instrument air
|
Pressure: 350 to 900 kPa, 500 to 900 kPa (with FPD)
350 to 900 kPa (Programmed temperature oven without cooler), 500 to 900 kPa (Programmed temperature oven with cooler) Maximum flowrate: Type 1: 140 L/min, 440 L/min (with FPD) (280 L/min typ. at Tamb=25℃) Type 2: 210 L/min, 510 L/min (with FPD) (340 L/min typ. at Tamb=25℃) Type 3: 280 L/min Type 4: 210 to 600 L/min (depends on specifications) Type 5: 210 L/min, 510 L/min (with 1 FPD)(340 L/min typ. at Tamb=25℃) , 870 L/min (with 2 FPDs)(550 L/min typ. at Tamb=25℃) Type 6: 70 L/min typ. (85 L/min max) |
|
Carrier gas
|
Types: H2, N2, He, or Ar
Purity: Measuring range from 0 to 50 ppm or more: 99.99% minimum (water: 10 ppm or less, organic components: 5 ppm or less) Measuring range from 0 to less than 50 ppm: 99.999% minimum (water: 5 ppm or less, organic components: 0.1 ppm or less) Pressure: H2: 500 ±20 kPa (to meet explosion-proof requirements, install a separate pressure reducing valve and supply at 500 kPa) Other than H2: 400 to 700 kPa Consumption: 60 to 300 mL/min per oven |